Measurement data transfer time – Atec Agilent-E4980A User Manual
Page 33
33
Measurement data transfer time
This table shows the measurement data transfer time under the following conditions.
The measurement data transfer time varies depending on measurement conditions and
computers.
Table 60. Measurement transfer time under the following conditions:
Host computer:
DELL OPTIPLEX GX260 Pentium 4 2.6 GHz
Display: ON
Impedance range mode:
AUTO (The overload has not been generated.)
OPEN/SHORT/LOAD compensation: OFF
Test signal voltage monitor: OFF
Table 61. Measurement data transfer time [ms]
using :FETC? command
using data buffer memory
Data
(one point measurement)
(list sweep measurement)
transfer
Comparator
Comparator
10
51
128
201
Interface format
ON
OFF
points points
points
points
ASCII 2 2 4
13
28
43
GPIB ASCII
Long 2
2
5 15
34
53
Binary 2 2 4
10
21
32
ASCII 2 2 3
8
16
23
USB ASCII
Long 2
2
4 9
19
28
Binary 2 2 3
5
9
13
ASCII 3 4 5
12
24
36
LAN ASCII
Long 3
3
5
13
29
44
Binary 3 3 5
9
18
26
DC bias test signal current (1.5 V/2.0 V):
Output current: Max. 20 mA
Option 001 (Power and DC Bias enhance):
DC bias voltage: DC bias voltage applied to DUT is given as:
Equation 18.
Vdut = Vb – 100 × Ib
Vdut [V] DC bias voltage
Vb [V]
DC bias setting voltage
Ib [A]
DC bias current
DC bias current: DC bias current applied to DUT is given as:
Equation 19.
Idut = Vb/(100 + Rdc)
Idut [A] DC bias current
Vb [V]
DC bias setting current
Rdc [Ω] DUT’s DC resistance