33

Measurement data transfer time

This table shows the measurement data transfer time under the following conditions.
The measurement data transfer time varies depending on measurement conditions and
computers.

Table 60. Measurement transfer time under the following conditions:

Host computer:

DELL OPTIPLEX GX260 Pentium 4 2.6 GHz

Display: ON

Impedance range mode:

AUTO (The overload has not been generated.)

OPEN/SHORT/LOAD compensation: OFF

Test signal voltage monitor: OFF

Table 61. Measurement data transfer time [ms]

using :FETC? command

using data buffer memory

Data

(one point measurement)

(list sweep measurement)

transfer

Comparator

Comparator

10

51

128

201

Interface format

ON

OFF

points points

points

points

ASCII 2 2 4

13

28

43

GPIB ASCII

Long 2

2

5 15

34

53

Binary 2 2 4

10

21

32

ASCII 2 2 3

8

16

23

USB ASCII

Long 2

2

4 9

19

28

Binary 2 2 3

5

9

13

ASCII 3 4 5

12

24

36

LAN ASCII

Long 3

3

5

13

29

44

Binary 3 3 5

9

18

26

DC bias test signal current (1.5 V/2.0 V):

Output current: Max. 20 mA

Option 001 (Power and DC Bias enhance):

DC bias voltage: DC bias voltage applied to DUT is given as:

Equation 18.

Vdut = Vb – 100 × Ib

Vdut [V] DC bias voltage
Vb [V]

DC bias setting voltage

Ib [A]

DC bias current

DC bias current: DC bias current applied to DUT is given as:

Equation 19.

Idut = Vb/(100 + Rdc)

Idut [A] DC bias current
Vb [V]

DC bias setting current

Rdc [Ω] DUT’s DC resistance