DC bias settling time

When DC bias is set to ON, add the following value to the settling time:

Table 66. DC bias settling time

Bias

Settling

time

1

Standard

Capacitance of DUT Ч 100 Ч log

e

(2/1.8 m) + 3 m

2

Option 001

Capacitance of DUT Ч 100 Ч log

e

(40/1.8 m) + 3 m

Figure 11. DC bias settling time

10 msec

100 msec

1 sec

10 sec

100 sec

1.

2.

1 µF

10 µF

100 µF

1 mF

10 mF

100 mF

DUT capacitance

Settling time