21

DC bias signal

Table 34. Test signal voltage

Range

–40 V to +40 V

Resolution

Setup resolution: 100 µV, effective

resolution: 330 µV ±(0 V - 5 V)

1 mV ±(5 V - 10 V)

2 mV ±(10 V - 20 V)

5 mV ±(20 V - 40 V)

Accuracy

test signal voltage ≤ 2 Vrms

0.1% + 2 mV (23

°

C ± 5

°

C)

(0.1% + 2 mV) x 4

(0 to 18

°

C or 28 to 55

°

C)

test signal voltage > 2 Vrms

0.1 % + 4 mV (23

°

C ± 5

°

C)

(0.1% + 4 mV) x 4

(0 to 18

°

C or 28 to 55

°

C)

Table 35. Test signal current

Range

–100 mA - 100 mA

Resolution

Setup resolution: 1 µA, effective

resolution: 3.3 µA ±(0 A - 50 mA)

10 µA ±(50 mA - 100 mA)

DC bias voltage level monitor Vdc

(0.5% of reading value + 60 mV) × Kt

When using Vdc-Idc measurement: (spec.)

When using level monitor: (typ.)

Kt

Temperature coefficient

DC bias current level monitor Idc

(A [%] of the measurement value + B [A]) × Kt

When using Vdc-Idc measurement: (spec.)

When using level monitor: (typ.)

A [%]

When the measurement time mode is SHORT: 2%

When the measurement time mode is MED or LONG: 1%

B [A]

given below

Kt

Temperature coefficient


When the measurement mode is SHORT, double the following value.