3 error codes list error codes list, Data errors, Tare errors – Ohaus E1M110 Explorer Balances User Manual
Page 43: Calibration errors, Rs232 errors, User errors, Over-under load errors, Checksum errors, 4 information messages
43
5.3 Error Codes List
Error Codes List
The following list describes the various error codes and which can appear on the display and the suggested remedy.
Data Errors
1.0
Transient error (hardware error, probably static discharge). If error persists, the balance must be
serviced.
1.4
Transducer data error (lost communications between logic boards).
Tare Errors
2.0
Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needs
recalibration.
Calibration Errors
3.0
Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.
RS232 Errors
4.4
RS232 buffer is full.
User Errors
7.0
User entry out of bounds.
7.2
Number outside of display capacity.
Over-Under Load Errors
8.0
Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balance
must be serviced.
8.1
Hardware error caused by an internal weight signal which is too high. Check load on the pan which may
be excessive. If error persists, the balance must be serviced.
8.2
Power-on load out of specification (LFT only)
8.3
Rated capacity exceeded. Remove excessive weight from pan.
8.4
Underload condition on balance. Check that the proper pan is installed.
8.5
Internal calibration weight internal sensor indicated its weight on the pan.
CheckSum Errors
9.0
Bad factory checksum. If error persists, have the balance serviced.
9.1
Bad factory checksum. If error persists, have the balance serviced.
9.2
Bad factory checksum. If error persists, have the balance serviced.
9.3
Bad factory checksum. If error persists, have the balance serviced.
9.4
Factory internal calibration data failed checksum. This failure will disable access to the InCAL
TM
feature (if
installed).
9.5
Factory calibration data failed checksum.
9.8
User calibration data failed checksum.
9.9
Factory temperature compensation data failed checksum.
5.4 Information Messages
CAL NOW
If InCAL
TM
(internal calibration) is installed. Message to recalibrate the balance. The
message will remain until calibrated.
WARM UP
The user tried to perform an internal or external calibration and this message will be
flashed in the 14 segment field. The balance requires a 30 minute warmup period. During
warmup the user can not perform calibration..
SAVED
This message is flashed when an item is changed in the menu and the new value is written
to the EEPROM.
LOCKED
This message is flashed when an item can not be changed in the menu because the menu
is locked and the Lockswitch is set locked.
LOW REF
The message is flashed in parts counting or percent when the calculated reference weight
is very low.
UNSTBLE
This message is flashed when the balance was unable to aquire stable data during internal
calibration.