5 percent weighing, Procedure – Ohaus E1M110 Explorer Balances User Manual
Page 22
22
3.5 Percent Weighing
Percent Weighing is
enabled only when Percent is turned ON. Percent weighing permits you to place a reference
load on the balance, then view other loads as a percentage of the reference. The load you place on the pan as a
reference may be displayed as any percentage you select from 5% to 100% (in 1% increments). One hundred percent
does not necessarily have to represent the reference load. Subsequent loads, displayed as a percentage of the
reference are limited only by the capacity of the balance. The default setting is Reference 100%. Refer to paragraph
4.8 to enable percent weighing.
NOTE: The PERCENT display (number of digits) is a
function of the accuracy of the balance and the size of
the reference weight. The display examples were with a
300g mass used with an 12 kilogram balance.
Procedure
• Remove the reference weight from the pan and re-
place it with another load. The second load is dis-
played as a percentage of the reference.
Balance displays reference weight for five seconds in
selected measuring unit, then displays the percentage.
• Press the Mode button.
• Press
or
button until PERCENT is dis-
played.
• Press Enter button, PUT>PAN 100% is displayed. If
a container is used, the balance can be tared at this
point.
• Put the reference load on the pan.
• Press
or
button and select reference
weight percentage (Percent Range 5 to 100). Hold
button down for fast change.
• Press Enter button to save setting, WORKING is
displayed ... calculating reference weight.