3.3.5 Calibration Test

Calibration test feature allows a check of a known calibration mass against the last stored calibration information in
the balance. Sample display illustrates an 12kg balance.

Procedure

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, CAL TYPE InCAL is displayed.

• Press

button to select CALTEST calibration,

CALtESt is displayed.

• Press Enter button, WORKING is displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.

• Display changes to PUT WT 10000 g.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds, WEIGH DIFF is displayed. The display
now indicates the actual difference in weight between
what value was just placed on the pan and the
previous weight value which was stored in the bal -
ance. After approximately 8 seconds,the display
returns to the WEIGH mode.

• Remove calibration test mass from the pan.