beautypg.com

6 parts counting, Procedure, Update – Ohaus E1M110 Explorer Balances User Manual

Page 23

background image

23

3.6 Parts Counting

Parts Counting is

enabled only when Count is turned ON in the Mode submenu and selected with the Mode button.

In the parts counting mode, the balance displays the quantity of parts you place on the pan. Since the balance
determines the quantity based on the average weight of a single part, all parts must be reasonably uniform in weight.

Procedure

• Press the Mode button.

• Press

or

button until COUNT is displayed.

• Press Enter button to save setting, PUT>PAN 100 PC

is displayed.

• Press >O/T< if taring is required.

• Press

or

button and select sample size.

Sample size is 5 to 1000 pieces.

• Place sample size on the pan.

• Press Enter button to continue, display indicates

WORKING.

Balance displays the reference weight of an individual
piece part for five seconds and then displays the total
number pieces on the pan.

• Remove the sample and place parts to be counted on

the pan. Balance displays number of pieces.

Update

Update is a function which permits placing additional
samples which are greater than the value of the original
sample but less than three times the value. This action
increases the accuracy of the measurement.

• Place sample on the pan which is at least one but not

more than three times the original sample size.

• Press Mode button, COUNT is displayed.

• Press Enter button, UPDATE is displayed.

• Press Enter button, WORKING is displayed then the

reference weight followed by the new sample size.