5 print, Procedure, Auto print feature – Ohaus E1M110 Explorer Balances User Manual
Page 29: Interval, Print stable data only, Print numeric data only
29
4.5 Print
The Print menu provides a number of options which can be turned ON or OFF. It contains eight submenus: Auto
Print, feature which includes selection of Off, Continuous, Interval and on Stability, Inter, specifies time interval for
automatic output of displayed data, Stable data-only feature, Numeric only or full display data for output, GLPCont,
GLPTare, Reference which prints reference weight value and Lock which enable you to program balance param-
eters and to lock the settings.
Procedure
• Press the Setup button, CAL is displayed.
• Press
or
button until PRINT is displayed.
• Press Enter button to continue.
• Press
or
button until either AUTOPRT,
INTER, STABLE, NUMERIC, GLPCONT, GLPTARE,
REFEREN, LOCK or EXIT is displayed.
• Press Enter button to continue.
• Press
or
button and select either menu
setting or ON or OFF.
• Press Enter button to save setting.
• Press
or
button to select next item or EXIT.
• Press Enter button to continue.
Auto Print Feature
When enabled, the Auto Print feature causes the balance to automatically output display data in one of three ways:
continuously, at user specified time intervals, or upon stability. Default settings are shown bold.
OFF
when set on turns off the auto print feature
Cont
when set on, outputs printed data continuously
Inter
provides a user specified printing interval
On Stb
provides printed data only when a stable reading is achieved
Interval
Can be set to provide a specified printing interval between 1 and 3600 seconds.
Print Stable Data Only
When set On , this feature permits only stable display data to be output. OFF is the default setting.
Print Numeric Data Only
When Numeric Data Only function is turned ON, this allows the balance to output numeric data only for RS232