3.3.1 Internal Calibration (InCAL

TM

)

On Explorer balances equipped with the InCal

TM

feature, calibration can be accomplished using an internal calibration

mass. When the balance requires calibration, a screen prompt of CAL NOW appears. Internal calibration can be
performed at any time providing the balance has warmed up to operating temperature.

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

Procedure

• Press Enter button, CAL TYPE InCAL is displayed.

• Press Enter button, INCAL is displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.

After a few seconds, CAL SET is displayed (the internal
weight is positioned and then removed), the display
then returns to WEIGH mode.

NOTE:

DO NOT DISTURB THE BALANCE DURING
CALIBRATION.