3.3.3 User Calibration

User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra-
tion feature, proceed as follows:

(FLASHING)

• Press the Setup button, CAL is displayed.

• Press Enter button, CAL TYPE is displayed.

• Press Enter button, CAL TYPE InCAL is dis -

played.

• Press

button twice to select USER calibration,

CAL TYPE USer is displayed.

• Press Enter button, the display indicates the last

calibration mass value which was entered with the

first digit flashing. (Sample illustrates 6000g).

• Press

and or

and enter the

desired mass value. This number must be at least
25% of the full span value.

• Press Enter button, WORKING is displayed.

NOTE: If a weight is left on the pan, the balance will
display CLR PAN (remove the weight from the pan).
The balance automatically resumes calibration.

• Display changes to PUT WT 6000g.

• Place specified calibration mass on pan.

• Press Enter button, WORKING is displayed. After a

few seconds CAL SET is displayed, the display then
returns to WEIGH mode.

User calibration is completed.

• Remove calibration mass from the pan.

Procedure