Rs232, 6 rs 232, 6 rs232 – Ohaus E1M110 Explorer Balances User Manual
Page 31: Baud rate, Parity, Data bits, Stop bits, Lock, Procedure
31
4.6 RS232
The RS232 menu provides communication parameters which can be set to accommodate external printers or com-
puters. It contains five submenus: Baud rate, Parity, Data, Stop bit and Lock ON or OFF which enable you to
program balance parameters and to lock the settings.
Baud Rate
This submenu is used to select the desired baud rate. There are five available baud rates to choose from: 300, 1200,
2400, 4800 and 9600. The default setting is 2400.
Parity
Parity can be set to Odd, Even or None. The default setting is None.
Data Bits
To set the number of data bits to 7 or 8. The default setting is 7.
Stop Bits
The number of stop bits can be set to 1 or 2. The default setting is 2.
Lock
Lock ON/OFF can only be changed when the hardware Lockswitch is set OFF/unlocked. A menu is locked when the
menu lock is set ON and the Lockswitch is ON. Lock when selected and turned on, locks all of the entries made
under the RS232 menu. In the locked condition, items may be looked at but not changed in the menu. When set off,
entries may be changed. OFF is the default setting.
Procedure
• Press the Setup button, CAL is displayed.
• Press
or
button until RS232 is displayed.
• Press Enter button to continue.
• Press
or
button until either BAUD, PARITY,
DATA, STOP, LOCK is displayed.
• Press Enter button to select.
• Press
or
button and select the desired
menu setting.
• Press Enter button to save setting.
• Press
or
button to select next item or EXIT.
• Press Enter button to continue.