beautypg.com

10 global (cont.) list, Reset, Version – Ohaus E1M110 Explorer Balances User Manual

Page 35: Lock

background image

35

LFT is Off
Mode Menu

WEIGH= On
PERCENT= Off
COUNT= Off
ANIMAL= Off

Lock Switch is Off
Menu Locks

RS232= Off
READOUT= Off
GLPSET= Off
MODE= Off
UNITS= Off
PRINT= Off
GLP Data = Off
CAL= Off
GLOBAL= Off
CUSTOM= Off

Enabled Units:

g
custm

C. Units:

1.000000 E0 x 1

4.10 Global (Cont.)

List

This submenu can be used to output a listing of current menu settings via the RS232 interface. When YES is selected,
all menu settings will be output either to an external printer or computer. To use this feature, your balance must be
connected to a computer or printer. The default setting is OFF.

EXPLORER xxxxxx-xxx
Sr #x.xx OS#x.xx G#x.xx

Time= 12hr 3:19:51 PM
Date= m/d/y 7/09/97
Function = Weigh
ReadOut Menu

Stb= 2 d
AZT= .5 d
Filter= 1

GLP Menu

Time/Date= On
Bal Id= On
User No.= On
Proj No.= On
DIFF= On
Name= On

Print Menu

Auto Print= Off
Interval= 7
Stable Print= On
NU= On
GLP Cont = Off
GLP on Tare = Off
Print Ref= On

RS232 = 2400: N: 7: 2

SAMPLE PRINTOUTS

The partial sample shown, indicates the status in the
menus.

Reset

Reset when set to yes will reset the balance to factory default settings. The default setting is NO.

Version

Displays software revision number for servicing purposes. This number is installed with the balance.

Lock

Lock when selected and set tp YES, locks all of the entries made under the Global menu. The default setting is NO.