3.3 Calibration (Cont.)

Calibration Menu Protection

NOTES:
• Calibration may be locked out to prevent unauthorized
personnel from changing calibration. If calibration has
been locked out, you can only access Internal Calibration
(InCAL

TM

) when installed and Cal Test.

• To lock out calibration menu, after calibration, refer to the
section titled Menu Lock-Out Protection.

Calibration Masses

Before beginning calibration, make sure masses are
available. If you begin calibration and realize calibration
masses are not available, exit the menu. The balance will
retain previously stored calibration data. Calibration
should be performed as necessary to ensure accurate
weighing. Masses required to perform the procedures are
listed in the adjacent table.

NOTE:
Any of the calibration modes can be terminated

at any

time by pressing either the Mode, Units or Setup buttons.

CALIBRATION MASSES

LINEARITY SPAN ONLY
CAPACITY MASSES

MASSES

12,000g

5,000g/10,000g

10,000g

22,000g

10,000g/20,000g

20,000g

32,000g

15,000g/30,000g

30,000g

It is recommened that masses must meet or exceed
ASTM Class 1 Tolerance. Calibration masses are
available as accessories.