Introduction, Description, Figure 2 – M&C TechGroup MP12 Operator's manual User Manual
Page 7: Performance characteristics mp06/12

7
Gas sampling and gas conditioning technology
6-1.1.10-ME
6
INTRODUCTION
Due to the implementation of resistant materials the pump type MP06/12 can be used in a wide range
of applications with aggressive and corrosive gases.
7
DESCRIPTION
The pump head and all wetted parts of the diaphragm pumps type MP06/12 are made of PTFE re-
spectively FKM. This grants a high resistance against aggressive and corrosive components in the
sample.
The pumps work 100% lubricant free which means that the sample remains analytical unaffected.
The special diaphragm and valve system guarantees a high level of reliability and a minimum of main-
tenance.
The pumps are available in 230V 50Hz.
The maximum flow capacity of the MP06 is 5,5 l/min. For higher flow up to maximum 11 l/min we
recommend the pump type MP12.
The range of the sample temperature during operation lies between
–20 °C and maximum +80 °C.
N O T E !
The diaphragm pumps series MP06/12 are not allowed for liquid medium.
Maximum temperature for medium during operating: -20°C .... +80°C. Before
use please check the compatibility of the material.
Figure 1
Performance characteristics MP06/12