Gas sampling and gas conditioning technology

9-3.15-ME

Safety handbook according to SIL


Contemplated devices

It was contemplated the PMA 50 EX with part no. 05A1000(a) and the following options:

-

Ventilation equipment and enclosure purging (part no. 05A9005 and 05A9000),

-

Measuring cell solvant resistant or with glass solder (part no. 90A0009 u. 90A0006)

-

Transmitter for measurements of gases from ex-zone 0

II 1/2 G EEx de [ia] IIC T5


Excluded are the

PMA 50 EX/P/PD… with pressure compensation, part no. 05A2500(a) as well

as the options 0-20mA outlet, part no. 05A2505(a) resp. zero suppression, part no. 05A9015.


Device description and safety function

The safety function of the device is the measurement of the oxygen concentration in the measu r-
ing cell which is provided as linear current signal 4-20mA. The status relay as transmission of a
summary fault indication is element of the safety function. Current signals <3,2mA and > 20,5mA
as well as an open status contact have to be evaluated as failure from a downstream device.

The SIL-qualification is valid for all adjustable measuring ranges. It applies for alarm of rising ox-
ygen concentration (dangerous fault: oxygen signal is to small) as well as for alarm of decreasing
oxygen concentration (dangerous fault: oxygen signal is too big).

For one channel and two channel operation of the oxygen analyser PMA 50 Ex the following parame-
ters have been determined.

Single channel 1oo1 Redundant 1oo2

Single channel 1oo1

Redundant

1oo2

Safety function

Mesurement of the oxygen

concentration

Mesurement of the oxygen

concentration

Dangerous error

Oxygen signal too low

Oxygen signal too big

Measuring range

Depending on the measuring task

Depending on the measuring task

Device type

B

(but without µP and software)

B

(but without µP and software)

Proof test interval

1 year

1 year

MTTR

24 hr

24 hr

SFF

91,3 %

96,4 %

HFT

0

1

0

1

SIL

–ability

2



2





factor

5 %

5 %

PFD

3,83 Ч 10

-4

1,93 Ч 10

-5

1,66 Ч 10

-4

8,34 Ч 10

-6

du

8,44 × 10

-8

(per hr)

3,52 × 10

-8

(per hr)

dd

4,62 × 10

-7

(per hr)

4,62 × 10

-7

(per hr)

su

4,25 × 10

-7

(per hr)

4,74 × 10

-7

(per hr)

sd

5,70 × 10

-11

(per hr)

5,70 × 10

-11

(per hr)