Zero calibration – M&C TechGroup PMA 50 EEX Operator's manual User Manual
Page 22
22
Gas sampling and gas conditioning technology
9-3.15-ME
18.1
ZERO CALIBRATION
The calibraiton of the zero point is to be effected with a gas free of oxygen, eg. Nitrogen (N
2
) 5.0.
The following steps have to be executed:
Connect the flexible PVC or Viton
tube with the bottle’s pressure reducer of the N
2
-zero gas
bottle;
W A R N I N G !
The pressure reducer should have a maximum outlet measuring
range of 0-1,5 bar abs. and should always be adjusted to a low
outlet pressure of max. 0,1 bar. This protects the measuring cell
against destruction due to high pressure!
First of all, open the bottle valve, then open the outlet valve of the pressure regulator;
The pressure regulator and the tubing have to be flushed for about 5 seconds;
Check adjusted control pr
essure and reduce to ≤ 0,1bar if necessary then shut the outlet valve
of the pressure regulator again;
Connect the free tube end of the zero gas bottle connection with the external flow meter or, if
available, with the external calibration valve;
Open the outlet valve of the pressure regulator slowly, in order to avoid pressure peaks;
Adjust the zero gas volume flow with the needle valve of the flow meter to max. 60 l/h. The
volume flow of the calibration gas should always be adapted to the sample gas volume flow;
N O T E !
Always calibrate at the gas flow used for the measurement too.
Wait approx. 30 seconds until stabilisation of the display;
Set the range switch to 0-1% O
2
;
If necessary, adjust the zero point to 0 % via the zero point potentiometer 12 (figure 1) (with a
screw driver);
Check the analogue output signals at 0% O
2
;
Ouput signal
Signal to be measured
0-1 V DC
0 V
0-20 mA
0 mA
4-20 mA
4 mA
N O T E !
Analysing a gas mixture the single components have to be checked ac-
cording to a possible cross sensitivity and to be considered for zero cali-
bration (see chapter 20).