7 input timing – IDK MSD-54 Series Command Guide User Manual
Page 49
MSD-54 Series Users Guide (Command Guide)
49
3.3.7 Input timing
@SHT
Set the total number of horizontal dots
Command
format
@SHT, ch, h_total
Return value
format
@SHT, ch, h_total
Parameter
ch
: Input channel
0 = All analog inputs, 6 = IN6 to 9 = IN9
h_total
: The total number of horizontal dots
400 to 4125 (Sampling clock: 13 MHz to 162 MHz)
[Default]: Differs depending on input signals
Example Send
Receive
@SHT,6,1344
@SHT,6,1344
Set the total number of horizontal dots of IN6
to 1344.
Terminated normally.
Send
Receive
@SHT,6,1344
@ERR,3
If no analog RGB or analog YPbPr signals are
input, an error is replied.
Related info.
8.6.1 The total number of horizontal dots
Note
This command is valid only if analog RGB or analog YPbPr signals are input.
@GHT
Get the total number of horizontal dots
Command
format
@GHT
Return value
format
@GHT, in_1, in_2, in_3, in_4, in_5, in_6, in_7, in_8, in_9
Parameter
in_1-9
: The total number of horizontal dots of each input channel
400 to 4125 (Sampling clock: 13 MHz to 162 MHz)
[Default]: Differs depending on input signals
Example Send
Receive
@GHT
@GHT,2200,2200,0,2640,
1344,1792,0,1792,0
Get the total number of horizontal dots.
0 is replied to channels without input
signals.
Related info.
8.6.1 The total number of horizontal dots