beautypg.com

2 current calibration (vxi plug&play demo), Figure 4-4. current calibration window, Current calibration (vxi plug&play demo) -8 – KEPCO KLR Series User Manual User Manual

Page 86: Current calibration window -8, R 4.4.2 for

background image

4-8

KLR 091313

4.4.2

CURRENT CALIBRATION (VXI plug&play DEMO)

During current calibration, the current, current readback and analog current readback are cali-
brated.

1. Click the CURRENT button on the Calibration Window to start current calibration; a pop-up

box appears instructing the operator to connect the shunt across the output terminals. Once
the shunt is connected, click OK; tthe Current Calibration Window (Figure 4-4) appears and
the status display on the VXI main panel reads CMAX.

FIGURE 4-4. CURRENT CALIBRATION WINDOW

2. Monitor DVM and click the “+” button to increase and the “–“ to decrease the output current

until the DVM reads as close as possible to maximum rated current (Calculate current as fol-
lows

:

I (Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).

3. Click the PRESS TO ACCEPT button to accept the value. The status display reads C_0

.

Click the “+” button to increase and the “–“ to decrease the output current until the DVM
reads as close as possible to minimum positive or zero current (Calculate current as follows:

(Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).

4. Click the PRESS TO ACCEPT button to accept the value. The status display reads CMAX.

Click the “+” button to increase and the “–“ to decrease the output current until the DVM
reads as close as possible to maximum rated current (Calculate current as follows:

I (Amperes) = DVM reading (Volts) / Shunt Resistance (Ohms).