beautypg.com

Figure 4-3. voltage calibration window, Voltage calibration window -7, E 4-3) – KEPCO KLR Series User Manual User Manual

Page 85

background image

KLR 091313

4-7

FIGURE 4-3. VOLTAGE CALIBRATION WINDOW

2. Click the Press to ACCEPT button to accept the value. The status display then reads

VMAX. Monitor

DVM and click the “+” button to increase and the “–“ to decrease the output

voltage until the DVM reads as close as possible to maximum rated voltage.

3. Click the Press to ACCEPT button to accept the value. The status display will not change.

At the calibration window, click the CURRENT Button to proceed to current calibration (see
PAR. 4.4.2)

.

4. Click the Press to ACCEPT button to accept the value; the status display now reads OUTV.

At the ANALOG I/O PORT J2, connect the DVM to VOLT_RBACK and GND as shown in
Figure 4-1. Click the “+” button to increase and the “–“ button to decrease the output voltage
until the DVM reads as close as possible to 10V.

5. Click the Press to ACCEPT button to accept the value. The status display will not change.

At the calibration window, click the CURRENT Button to proceed to current calibration (see
PAR. 4.4.2)

.