To configure the final scan parameters, To configure the final scan parameters -5 – RIGOL DSA875 User Manual
Page 43

Chapter 6 To Set the Scan Parameters
RIGOL
User’s Guide for EMI Test System
6-5
Note: When adding the scan sub-ranges, the sub-ranges cannot overlap.
Otherwise, "Subrange is repeat. Please change the range." will be displayed.
2. Delete data points
Select a row of data in the scan list and click Delete to delete the scan
data in the specified row.
Click Clear to delete all the data in the scan list.
3. Save the data list
Click Save to save the scan data edited in csv format.
4. Load a scan list
Click Load to load a stored scan data csv file.
To Configure the Final Scan Parameters
Set the following parameters in Final Scan Configure in the scan parameter
editing interface (Figure 6-1).
1. Final scan detector
Click the dropdown arrow in the Final-Det box and the menu as shown in the
figure below is displayed. The detector modes available include Qpeak,
Average and RMS.
2. Final scan resolution bandwidth
Click the Final-RBW box and the menu as shown in the figure below is
displayed. You can set the current scan resolution bandwidth to 10Hz (-3dB),
30Hz (-3dB), 100Hz (-3dB), 200Hz (-6dB), 300Hz (-3dB), 1kHz (-3dB), 3kHz
(-3dB), 9kHz (-6dB), 10kHz (-3dB), 30kHz (-3dB), 100kHz (-3dB), 120kHz
(-6dB), 300kHz (-3dB) or 1MHz (-3dB).
When the final scan detector (Final-Det) is set to "Qpeak", only 200Hz
(-6dB), 9kHz (-6dB) and 120kHz (-6dB) are available.