To configure the segment scan parameters, To set the scan sub-range parameters, To configure the segment scan parameters -3 – RIGOL DSA875 User Manual
Page 41: To set the scan sub-range parameters -3

Chapter 6 To Set the Scan Parameters
RIGOL
User’s Guide for EMI Test System
6-3
To Configure the Segment Scan Parameters
To Set the Scan Sub-range Parameters
You can set the following parameters in SubRange in the scan parameter editing
interface (Figure 6-1).
1. Sub-range start frequency
Click the Range Start(Hz) box and input the desired start frequency of the
sub-range. The default unit is Hz.
2. Sub-range stop frequency
Click the Range Stop(Hz) box and input the desired stop frequency of the
sub-range. The default unit is Hz.
3. Scan mode
Click the dropdown arrow in the Sweep Mode box to set the scan mode
within the scan sub-range currently selected to Auto or Manual.
When Auto is selected, the Step Size(Hz) and Meas Time(ms) boxes
are grayed out and disabled. When the pre-scan resolution bandwidth
(Pre-RBW) is set to different value, Step Size(Hz) uses the
corresponding default value.
When Manual is selected, you can set the Step Size(Hz) and Meas
Time(ms) manually. For the setting methods, refer to the setting of
4. Pre-scan resolution bandwidth
Click the Pre-RBW box and the menu as shown below is displayed. You can
set the current scan resolution bandwidth to 10Hz (-3dB), 30Hz (-3dB), 100Hz
(-3dB), 200Hz (-6dB), 300Hz (-3dB), 1kHz (-3dB), 3kHz (-3dB), 9kHz (-6dB),
10kHz (-3dB), 30kHz (-3dB), 100kHz (-3dB), 120kHz (-6dB), 300kHz (-3dB) or
1MHz (-3dB).
When the pre-scan detector (Pre-Det) is set to "Qpeak", only 200Hz (-6dB),
9kHz (-6dB) and 120kHz (-6dB) are available.