II - 5 Measured Value Output

102

Technical Information

POSITIP 855

Fig. 45: Time diagram for measured value

output with Ctrl B

II - 5

Measured Value Output

POSITIP can output measured values over the data interface.

Starting measured value output

There are three ways to start measured value output:

• Transmit control character to the data interface

• Send signal to switching input

• Signal from edge finder
The delay between the latch signal and measured value output

depends on the selected signal.

Transit time of encoder signals
After approximately 4 µs the encoder signals are present in a buffer

which is interrogated by the internal latch signal. The measured

value that is output is therefore the value that existed approximately

4 µs prior to the internal latch.

Starting measured value output with Ctrl B

t

1

: Delay between Ctrl B command and internal latch

t

1

£ 0.5 ms

t

2

: Delay between internal latch and measured value output

t

2

£ 30 ms + (5 ms · N)

N = number of rotary axes with Deg/Min/Sec display

t

3

: Time between end of data output and next latch with

Ctrl B

t

3

³ 0 ms

t

D

: Duration of measured value output

The duration of measured value output (t

D

) depends on:

• The selected baud rate (BR)

• The number of axes (M)

• The number of blank lines (L)

t

D

=

[s]

C t r l B

C t r l B

t

1

t

3

t

2

t

D

R X D

T X D

BR

176 · M + L · 11