Ii - 5 measured value output, Starting measured value output – HEIDENHAIN PT 855 for Milling User Manual
Page 104

II - 5 Measured Value Output
102
Technical Information
POSITIP 855
Fig. 45: Time diagram for measured value
output with Ctrl B
II - 5
Measured Value Output
POSITIP can output measured values over the data interface.
Starting measured value output
There are three ways to start measured value output:
Transmit control character to the data interface
Send signal to switching input
Signal from edge finder
The delay between the latch signal and measured value output
depends on the selected signal.
Transit time of encoder signals
After approximately 4 µs the encoder signals are present in a buffer
which is interrogated by the internal latch signal. The measured
value that is output is therefore the value that existed approximately
4 µs prior to the internal latch.
Starting measured value output with Ctrl B
t
1
: Delay between Ctrl B command and internal latch
t
1
£ 0.5 ms
t
2
: Delay between internal latch and measured value output
t
2
£ 30 ms + (5 ms · N)
N = number of rotary axes with Deg/Min/Sec display
t
3
: Time between end of data output and next latch with
Ctrl B
t
3
³ 0 ms
t
D
: Duration of measured value output
The duration of measured value output (t
D
) depends on:
The selected baud rate (BR)
The number of axes (M)
The number of blank lines (L)
t
D
=
[s]
C t r l B
C t r l B
t
1
t
3
t
2
t
D
R X D
T X D
BR
176 · M + L · 11