Step 3
– Configure Boards
DaqBoard/1000 and /2000 Series Boards have no jumpers or switches to set. Configuration is performed,
in its entirety, through software. Refer to the following figure and steps to complete the configuration.
The numbers in the figure correspond to the numbered steps immediately following the figure.
Accessing the DaqBoard/2000 Properties Tab
Note: Images are similar for DaqBoard/1000 Series.
1.
Run the Daq Configuration control panel applet. Navigation from the desktop to the applet is as follows:
Start
⇒ Settings ⇒ Control Panel ⇒ Daq*Configuration (double-click)
2.
Double-click on the Device Inventory’s DaqBoard2K0 or DaqBoard1K0 icon, as applicable. The DaqBoard’s
Properties tab will appear. If the DaqBoard icon is not present, skip to the upcoming Using ‘Add Device’
section.
3.
Enter a “Device Name” in the text box, or use the default, e.g., DaqBoard2K0. The Name is for identifying the
specific DaqBoard, but actually refers to the PCI slot.
4.
Verify that the “Device Type” shows the correct board, e.g., “DaqBoard/1000, DaqBoard/2001, etc.”
Note that available device types can be viewed via the pull-down list.
5.
Confirm that the DaqBoard’s text box shows a Bus #, Slot #, and Serial Number.
If this text box is empty, use its pull-down list and select the serial number that matches the one for your board.
1033-0940, rev 8.0
898195
DaqBoard/1000 and /2000 Series Installation Guide IG-5