Analog output calibration – Flowserve Logix MD+ ValveSight User Manual
Page 272

ValveSight™ Diagnostics DTM Manual for Logix MD+ Positioner with HART®
FCD-
LGENSF0014-00
© Flowserve Corporation
272
 
 
Analog Output Calibration
If the Multi-Function Card (MFC) is configured as an Analog Output, this page allows you to 
calibrate the the configuration of the MFC Card to function as an Analog Output. 
 
Status Area 
The Active Indicator area shows the status of the most relevant active indicator. The color 
of the "LED light" corresponds to the Active Indicator and the first color of the blink code 
sequence on the positioner. Generally green indicates no immediate issues. Yellow 
indicates a developing issue. Red indicates the ability to control may be compromised. A 
detailed list of the indicators is given in the 
Alarm Congfiguraion - Alarm Annunicator
page.
The Command Source field indicates weather the positioner is being controlled by digital or 
an analog (4-20 mA) command source. 
 
Calibration Information 
This area displays information about the last analog output calibration. 
