How to enter calibration mode of ch3 : press the, Rotary & numerical key “3” simultaneously -4 – AMETEK XBT Series User Manual

Page 58

background image

Calibration procedure

• First, you can read the screen of the current calibration mode of CH2,

XBT32-3FTP shall output one low current. Please read the actual reading from
your electric meter and input the value to XBT32-3FTP. Then press the rotary to
confirm the data input. Second, XBT32-3FTP shall output one high current.
Please check the reading of electric meter and input the actual value.Once the
data input is confirmed, the current calibration procedure is finished.

6.1.3 How to enter calibration mode of CH3 : Press the rotary &

numerical key “3” simultaneously.

Wiring to calibrate CH3 (voltage mode)