Remote Interface Protocol and Package Mode

How to read back serial number:

MEM:SERIAL?
MEMORY:SER?

How to modify machine ID (address):

CONTROL:ADDR 23

==> modify id to 23

CONT:ADDRESS 09

==> modify id to 9

How to check password:

CONT:PASSWORD 123456

==> enter password to verify

How to enter or leave the monitor mode

(not supported over GPIB or Ethernet LAN):

CONT:MONITOR ON

==> enter the monitor mode

CONTROL:MONI OFF

==> exit the monitor mode

How to change channel:

CONT:CHANNEL 1

==> change to channel 1

CONT:CH 1

==> change to channel 1

CONTROL:CH 2

==> change to channel 2

How to enter or leave lock status:

CONT:LOCK ON

==> enter lock mode

CONTROL:LOCK OFF

==> exit lock mode

How to enter or leave hotkey status:

CONT:HOTK ON

==> enter hotkey mode

CONTROL:HOTKEY OFF

==> exit hotkey mode

How to measure current:

MEASURE:CURR1? ==>

read

back current 1 result

MEAS:CURRENT3? ==>

read

back current 3 result

MEAS:CURR3? ==>

read

back current 3 result

CURR1?

==> read back current 1 result

CURRENT2?

==> read back current 2 result

5-12