6.

CALIBRATION PROCEDURE

6.1 Requirement of calibration instrument

* Electric meter (6 ½- digital).
* Please note if the current shunt of electric meter is too large, the current
measurement will be not correct.
* The current of the third channel can reach 5A. Therefore, please make you use
the current mode at 5A (or larger). Or it will damage your meter.

6.1.1 How to enter calibration mode of CH1 : Press the rotary &

numerical key “1” simultaneously.

Wiring to calibrate CH1 (voltage mode)

• First, you can read the screen of the voltage calibration mode of CH1,

XBT32-3FTP shall output one low voltage. Please read the actual reading from
your electric meter and input the value to XBT32-3FTP. Then press the rotary to
confirm the data input. Second, XBT32-3FTP shall output one high voltage.
Please check the reading of electric meter and input the actual value.Once the


M370187-01

Rev J

6-1