AMETEK XBT Series User Manual
Page 47

Remote Interface Protocol and Package Mode
IOUT1?
==> read back current 1 result
IOUT2?
==> read back current 2 result
How to measure voltage:
MEASURE:VOLT1?
==> read back voltage 1 result
MEAS:VOLTAGE3?
==> read back voltage 3 result
MEAS:VOLT3?
==> read back voltage 3 result
VOUT1?
==> read back voltage 1 result
VOUT2?
==> read back voltage 2 result
How to measure power:
MEASURE:POW1?
==> read back power for ch1
MEAS:POWER3?
==> read back power for ch3
MEAS:POW3?
==> read back power for ch3
How to measure resistance:
MEASURE:RES1?
==> read back resistance for ch1
MEAS:RESISTANCE3? ==>
read
back resistance for ch3
MEAS:RES3?
==> read back resistance for ch3
How to set output voltage:
SOUR:VOLTAGE2 12
==> set output voltage to 12V for channel 2.
SOURCE:VOLT1 30
==> set voltage to 30V for channel 1.
VOLT3 10
==> set voltage to 10V for channel 3.
VOLTAGE3 5
==> set voltage to 5V for channel 3.
VSET2 15
==> set voltage to 15V for channel 2.
How to set output current:
SOUR:CURRENT2 1
==> set current to 1A for channel 2.
SOURCE:CURR1 3
==> set current to 3A for channel 1.
CURR3 1.2
==> set current to 1.2A for channel 3.
CURRENT3 5
==> set current to 5A for channel 3.
ISET2 1.5
==> set current to 1.5A for channel 2.
How to define over voltage protection:
M370187-01 Rev J
5-13