3 power-line voltage, 4 fuse, 5 connections – AMETEK XBT Series User Manual
Page 18: 6 warming up

Installation and Maintenance
2-2
M370187-01 Rev J
2.3 Power-Line Voltage
The power supply uses AC power 115V/230V 50Hz/60Hz. Before plugging in the
power cord, make sure the POWER switch (see Section 3.1) is in the OFF
position and the voltage input selection switch on the bottom of the unit is the
same as the required voltage.
Caution
Exceeding the maximum rated AC input voltage could result
in damage to the unit.
2.4 Fuse
There is one fuse installed in the rear panel. If and when it becomes necessary to
replace the fuse, first turn off the power and disconnect the AC power cord and
all other connections to the power supply. Open the fuse cover, located on the
rear panel below the AC socket, using a flat-head screwdriver or by holder from
under the AC socket.
Mark Range Fuse
Max
Power
115 104V~126V
T5A
/250V 500VA
230 207V~253V
T2.5A
/250V
500VA
50/60 Hz
Caution
For continuous protection against fire hazard, replace fuse
only with the same type and rating specified.
2.5 Connections
Of the interface connections available, RS-232 and USB are standard interfaces,
and the GPIB and Ethernet interfaces are options.
2.6 Warming Up