Page 63

background image

Sorensen DCS-E Series 3kW Supplies

Advanced Operation

M362295-01 Rev C

4-19

4.5.3 Shutdown

Application - Contact Closure

An external relay, whether normally open or normally closed, may be used to activate the
Shutdown circuit. Either positive or negative logic may be used.

To activate the Shutdown function using an external relay:

1. Turn off the power supply and disconnect the AC power source. Allow five (5) minutes to

elapse to dissipate stored energy before making connections or resetting switches.
Refer to Section 4.2 for instructions for resetting switches or making J3 connections.

2. Connect one side of the external relay to pin 15 (+15 Vdc Auxiliary Output) on connector

J3. Connect the other side of the relay to pin 14 (TTL Shutdown Input). Also connect
pin 2 (Shutdown Return) to pin 6 (Auxiliary Supply Ground). See Figure 4-18 through
Figure 4-21
.

3. Set internal switch SW1-8 to select the desired circuit logic as shown in the table below.

Relay

Switch SW1-8 Setting

Relay Coil State

Output

Normally Open Relay

OPEN (Positive Logic)

Energized

OFF

De-energized

ON

CLOSED (Negative Logic)

Energized

ON

De-energized

OFF

Normally Closed Relay

OPEN (Positive Logic)

Energized

ON

De-energized

OFF

CLOSED (Negative Logic)

Energized

OFF

De-energized

ON

The red S/D (Shutdown) LED on the front panel lights up when the Shutdown circuit is
activated.

Figure 4-18 Using Shutdown with Contact Closure

of a Normally OPEN Relay (Positive Logic)