Advanced Operation

Sorensen DCS-E Series 3kW Supplies

4-10

M362295-01 Rev C

Figure 4-5 Programming Output Voltage with a 0-10 Vdc Source

WARNING
Exercise caution when resetting internal jumpers. Review Section 2.2 Safety and
follow procedures for removing the cover and resetting jumpers in Section 4.2.3
Resetting Jumpers and Switch.

P

ROGRAMMING

O

UTPUT

V

OLTAGE WITH A

0-1

M

A S

OURCE

1. Remove the jumper between pins 20 and 21 of connector J3.

2. Connect the external programming source between pin 9 (voltage programming

input/positive) and pin 12 (return) of connector J3.

Varying the current source from 0 to 1 mA will vary the output voltage from 0 to 100% of the
model rating. Adjusting the front panel current limit control sets the output current limit.

Figure 4-6 Programming Output Voltage with a 0-1 mA Source