• Inspect the electrode for scratches or cracks. If any is present, replace the

electrode.

• Cleaning procedure: clean frequently the probe by soaking it for 1 minute in

HI70670

HI70670

HI70670

HI70670

HI70670 or HI70671

HI70671

HI70671

HI70671

HI70671 cleaning solution. After cleaning soak the electrode in

HI70300

HI70300

HI70300

HI70300

HI70300 storage solution before taking measurements.

Maintenance of EC Probe
• After every series of measurements, rinse the probe with tap water.
• If a more thorough cleaning is required, clean the probe with the supplied

brush or a nonabrasive detergent.

Notes
• After a cleaning procedure, always recalibrate the system.
• Grease the O-rings before installing back the sensors.