Hanna Instruments HI 98280 User Manual

Page 51

background image

51

7.2 pH

• From the “GLP data” menu, select the “pH” option and press .
• All data regarding last pH calibration appear:

offset, acidic slope, basic slope, used buffers,
time and date of the procedure.

• Use the arrow keys to scroll the stored data of

last 5 calibrations.

• Press ESC to return to the “GLP data” menu.

Note

Note

Note

Note

Note

The “C” label near the buffer value indicates a custom point, while the
“H” indicates HANNA standard value.
If a quick calibration was performed, the buffer values are replaced
with the “Quick calibration” indication.
If calibration was cleared, the offset and slope values are the default
ones, and the message ”Old calibrations cleared” appears.
If no pH calibration has been performed, a warning message advise
the user. Press to return to the previous screen.

7.3 DISSOLVED OXYGEN

• From “GLP data” menu select the “Dissolved

oxygen” option and press .

• All data regarding last DO calibration appear:

calibrated points, % saturation or concentra-
tion, time and date.

• Use the arrow keys to scroll the last 5 memo-

rized calibrations.

• GLP calibration data for DO include 3 options: 2-point % DO calibration,

single point % DO calibration and concentration DO calibration.

Notes

Notes

Notes

Notes

Notes

The “C” label near the calibration point indicates a custom point, while
the “H” means HANNA standard value.
When the % DO range is calibrated, also the DO concentration range
is calibrated, and vice versa.
If no DO calibration has been performed, a warning message advises
the user. Press to return to the previous screen.