• After confirmation, the following messages are displayed: “Storing data on

probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.

• Press OK to return to the “Calibration” menu.
• To return to the main menu, press ESC repeatedly.
• To restore the factory calibration data, select the corresponding option in the

“ORP calibration” menu and then press .

4.7 TEMPERATURE CALIBRATION

The meter is factory calibrated for temperature readings. If necessary, tempera-
ture calibration may be performed as explained below.

• Select “Temperature” option from the “Calibration” menu and press to

enter the temperature calibration mode.

• Insert the probe in a thermostatic bath.
• Select the measure unit (°C, °F or K) and in-

sert the bath temperature value (measured with
a reference thermometer).

• When the reading in stable, “Ready” appears

on the LCD.

• Press to store the calibration point.
• After confirmation, the following messages are

displayed: “Storing data on probe, please
wait...”, “Updating GLP data, please wait...”
and “Calibration completed”.

• Press to return to the “Calibration” menu.
• To return to the main menu, press ESC repeatedly.

Note

Note

Note

Note

Note

The meter allows a maximum difference
of ±2°C between the current reading
and the set value. If this condition is
not satisfied, the display shows the
warning message “Max +/-2°C is al-
lowed”.