Hanna Instruments HI 98280 User Manual
Page 52

52
7.4 CONDUCTIVITY
• From “GLP data” menu select the “Conductiv-
ity” option and press
to view data for conductivity, actual conductiv-
ity and salinity calibrations.
• All information regarding last conductivity calibration appears: calibrated point,
cell constant value, calibration type (conductivity, actual conductivity or salin-
ity), time and date.
• Use the arrow keys to scroll the last 5 memorized calibrations.
• For conductivity calibration the following screens are available: conductivity,
actual conductivity, salinity.
Notes
Notes
Notes
Notes
Notes
The “C” letter near the conductivity calibration indicates a custom point,
while the “H” means HANNA standard value.
If no conductivity calibration has been performed, a warning message
advises the user. Press
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.
7.5 ATMOSPHERIC PRESSURE
• From “GLP data” menu select “Atmospheric pressure” option and press
• All information about last atmospheric pressure calibration appears: custom
calibration point, time and date.
• If factory calibration was restored, the display
shows the message “Factory calibr. restored”.
• Use the arrow keys to scroll the last 5 memo-
rized calibrations.
Notes
Notes
Notes
Notes
Notes
If no atmospheric pressure calibration has been performed, a warning
message advises the user. Press
If the selected calibration is a factory calibration, the meter shows the
message “Factory calibration”.