Hanna Instruments HI 98280 User Manual
Page 32

32
• Press
other buffer using the arrow keys.
• After the first calibration point is confirmed, immerse the probe in the second
buffer solution and stir gently.
• When the reading is stable and close to the selected buffer, the display shows
the “Ready” message.
• Press
• After the second calibration point is confirmed, immerse the probe in the third
buffer solution, stir gently and wait for stable reading.
• When the calibration is completed, the display shows the following messages:
“Storing data on probe, please wait...”, “Updating GLP data, please wait ...”
and “Calibration completed”.
• Press
• To return to the main menu, press ESC repeatedly.
Note
Note
Note
Note
Note
The pH calibration mode can be quitted at any time, by pressing the
ESC key.
Custom buffer calibration
HI9828 also allows a single point procedure to
calibrate with a custom buffer value.
• To select this option press first
then
ing for stable reading.
• A text box window will appear for inserting the desired custom value. The valid
range for custom buffer is from 0.00 to 14.00 pH.