Hanna Instruments HI 98280 User Manual

Page 14

background image

14

1.6 SPECIFICATIONS

TEMPERATURE
Range

Range

Range

Range

Range

-5.00 to 55.00 °C;

23.00 to 131.00 °F; 268.15 to 328.15 K

Resolution

Resolution

Resolution

Resolution

Resolution

0.01 °C; 0.01 °F; 0.01 K

Accuracy

Accuracy

Accuracy

Accuracy

Accuracy

± 0.15 °C; ± 0.27 °F; ±0.15 K

Calibration

Calibration

Calibration

Calibration

Calibration

Automatic at 1 custom point

p H

p H

p H

p H

p H

Range

Range

Range

Range

Range

0.00 to 14.00 pH; ± 600.0 mV

Resolution

Resolution

Resolution

Resolution

Resolution

0.01 pH; 0.1 mV

Accuracy

Accuracy

Accuracy

Accuracy

Accuracy

± 0.02 pH; ± 0.5 mV

Calibration

Calibration

Calibration

Calibration

Calibration

Automatic 1, 2 or 3 points with 5 memorized

standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01)

or 1 custom buffer

ORP

Range

Range

Range

Range

Range

± 2000.0 mV

Resolution

Resolution

Resolution

Resolution

Resolution

0.1 mV

Accuracy

Accuracy

Accuracy

Accuracy

Accuracy

± 1.0 mV

Calibration

Calibration

Calibration

Calibration

Calibration

Automatic at 1 custom point

DISSOLVED OXYGEN

Range

Range

Range

Range

Range

0.0 to 500.0 %

0.00 to 50.00 mg/L

Resolution

Resolution

Resolution

Resolution

Resolution

0.1 %

0.01 mg/L

Accuracy

Accuracy

Accuracy

Accuracy

Accuracy

0.0 to 300.0 %: ± 1.5 % of reading

or ± 1.0% whichever is greater;

300.0 to 500.0 %: ± 3% of reading

0.00 to 30.00 mg/L: ± 1.5 % of reading

or 0.10 mg/L whichever is greater;

30.00 mg/L to 50.00 mg/L: ± 3% of reading

Calibration

Calibration

Calibration

Calibration

Calibration

Automatic 1 or 2 points at 0, 100 % or 1 custom point