beautypg.com

Hanna Instruments HI 98280 User Manual

Page 36

background image

36

The 3 options are correlated, so that each of the procedures will calibrate all the
3 ranges.

Note

Note

Note

Note

Note

For correct EC readings, calibration should be performed using a stan-
dard solution with a conductivity value close to the sample to be mea-
sured.

After choosing the “Conductivity calibration” mode from the “Calibration” menu,
select the calibration type with the arrow keys and then press .

Conductivity
• Select the “Conductivity” option and press to confirm.
• Fill a beaker with a standard conductivity solution (see “Accessories” section for

choosing the proper HANNA solution).

• Immerse the probe in the solution and wait for stable reading. The probe sleeve

must be inserted.

• The main display shows the actual reading, while the secondary level displays

the standard value.

• To change the standard value, press and the list of available

standard values is displayed: 0 µS/cm, 84 µS/cm, 1413 µS/cm, 5.00 mS/cm,
12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm.

• Press to insert a custom value (temperature compensated value).

Choose to select the desired resolution.

• When the reading is stable, press to store the calibration data.
• After confirmation, the following messages are displayed: “Storing data on

probe, please wait...”, “Updating GLP data, please wait...” and “Calibration
completed”.

• Press to return to the “Calibration” menu.
• To return to the main menu, press ESC repeatedly.