CONDUCTIVITY
Range
Range
Range
Range
Range
0.000 to 200.000 mS/cm
(actual EC up to 400 mS/cm)
Resolution
Resolution
Resolution
Resolution
Resolution
Manual
1 µS/cm; 0.001 mS/cm; 0.01 mS/cm; 0.1 mS/cm; 1 mS/cm
Automatic
1 µS/cm from 0 to 9999 µS/cm
0.01 mS/cm from 10.00 to 99.99 mS/cm
0.1 mS/cm from 100.0 to 400.0 mS/cm
Automatic mS/cm
0.001 mS/cm from 0.000 to 9.999 mS/cm
0.01 mS/cm from 10.00 to 99.99 mS/cm
0.1 mS/cm from 100.0 to 400.0 mS/cm
Accuracy
Accuracy
Accuracy
Accuracy
Accuracy
±1 % of reading or ±1 µS/cm whichever is greater
Calibration
Calibration
Calibration
Calibration
Calibration
Automatic at 1 point with 6 memorized standards
(84 µS/cm, 1413 µS/cm, 5.00 mS/cm, 12.88 mS/cm,
80.0 mS/cm, 111.8 mS/cm) or custom point
RESISTIVITY
Range
Range
Range
Range
Range
0 to 999999
Ω
·cm;
(depending on measurement setup)
0 to 1000.0 k
Ω
·cm;
0 to 1.0000 M
Ω
·cm
Resolution
Resolution
Resolution
Resolution
Resolution
Depending on resistivity reading
Calibration
Calibration
Calibration
Calibration
Calibration
Based on conductivity or salinity calibration
TDS
Range
Range
Range
Range
Range
0 to 400000 mg/L or ppm;
(the maximum value depends on the TDS factor)
Resolution
Resolution
Resolution
Resolution
Resolution
Manual
1 mg/L (ppm); 0.001 g/L (ppt);
0.01 g/L (ppt); 0.1 g/L (ppt); 1 g/L (ppt)
Automatic
1 mg/L (ppm) from 0 to 9999 mg/L (ppm)
0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt)
0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt)
Automatic g/L (ppt)
0.001 g/L (ppt) from 0.000 to 9.999 g/L (ppt)
0.01 g/L (ppt) from 10.00 to 99.99 g/L (ppt)
0.1 g/L (ppt) from 100.0 to 400.0 g/L (ppt)
Accuracy
Accuracy
Accuracy
Accuracy
Accuracy
±1 % of reading or ±1 mg/L (ppm) whichever is greater
Calibration
Calibration
Calibration
Calibration
Calibration
Based on conductivity or salinity calibration