SCPI-Based Errors

260

IQS-600

–363

“Input buffer overrun” Software or hardware input buffer on serial port

overflows with data caused by improper or

nonexistent pacing.

–365

“Time out error”

This is a generic device-dependent error.

–400

“Query error”

This is the generic query error for devices that

cannot detect more specific errors. This code

indicates only that a Query Error as defined in IEEE

488.2, 11.5.1.1.7 and 6.3 has occurred.

–410

“Query INTERRUPTED” Indicates that a condition causing an

INTERRUPTED Query error occurred (see IEEE

488.2, 6.3.2.3); for example, a query followed by

DAB or GET before a response was completely

sent.

–420

“Query

UNTERMINATED”

Indicates that a condition causing an

UNTERMINATED Query error occurred (see IEEE

488.2, 6.3.2.2); for example, the device was

addressed to talk and an incomplete program

message was received.

–430

“Query DEADLOCKED” Indicates that a condition causing an

DEADLOCKED Query error occurred (see IEEE

488.2, 6.3.1.7); for example, both input and output
buffers are full and the device cannot continue.

–440

“Query

UNTERMINATED after

indefinite response”

Indicates that a query was received in the same

program message after an query requesting an

indefinite response was executed (see IEEE 488.2,

6.5.7.5).

–500

“Power on”

The instrument has detected an off to on transition

in its power supply.

–600

“User request”

The instrument has detected the activation of a

user request local control

Error

Number

Description

Probable Cause