EXFO IQS-600 Integrated Qualification System User Manual
Page 183

Using IQS Products in an Automated Test Environment
Integrated Qualification System
175
Managing the States of Your IQS-600 Integrated Qualification System
Note: See the appendix on COM properties and events for more information.
Note: You can also return to local state by turning the device off and on again.
However, this operation will cause the controller to lose control of the
system (no more local lockout state). All settings made on the device by the
controller will be lost (reset of device configuration when turning on).
4
The local lockout interface message (LLO) is
set to True.
The LockoutState property to
True.
5
The Clear Lockout/Set Local message is
received from the controller (the REM bus line
is set to False).
The LockoutState property to
False.
6
The device is addressed to listen (MLA).
The RemoteState property to
True.
7
The Clear Lockout/Set Local message is
received from the controller (the REM bus line
is set to False).
-----
8
The go to local interface message (GTL) is
received.
The RemoteState property to
False.
9
-----
The LockoutState property to
False.
No.
Possible in GPIB if:
Possible in ActiveX by setting: