beautypg.com

Stb and sre – EXFO IQS-600 Integrated Qualification System User Manual

Page 188

background image

Using IQS Products in an Automated Test Environment

180

IQS-600

Standard Status Data Structure

The following table presents a summary of the possible operations on ESR
and ESE registers.

STB and SRE

The status byte register and service request enable register information is
presented in the following table.

Register

Read

Write

Clear

ESR

Use *ESR?. Impossible

to write.

Use *CLS.

Read the register.

ESE

Use *ESE?. Use *ESE.

Use *ESE with a value equal to 0.

At IQS Manager startup, the

register is set to 0.

Use *ESE with a value equal to 0.

Bits

Mnemonics

Bit Value

7

Not Used (N.U.)

0

6

Master Summary Status (MSS)/ Service

Request (RQS)

64

5

Event Summary Bit (ESB)

32

4

Message Available (MAV)

16

3

Not Used (N.U.)

0

2

Error Available (EAV)

4

1

Not Used (N.U.)

0

0

Not Used (N.U.)

0