EXFO IQS-600 Integrated Qualification System User Manual
Page 223

Troubleshooting
Integrated Qualification System
215
Solving Common Problems
You receive an error message
indicating that you must insert
your high-speed module into
a high-speed unit even
though the module is already
inserted into a high-speed
unit.
At least one standard
unit is connected
before the high-speed
unit housing the
high-speed module.
The units must be
chained in a specific
order. Ensure that all
high-speed units are
connected before all
standard units.
The high-speed
module is not
inserted into one of
the first three
high-speed units.
Ensure that the
high-speed module is
inserted in one of the first
three high-speed units.
A USB device is not
functioning.
The USB device was not
recognized by the system.
Disconnect, then reconnect
the USB device.
SCPI commands are not
displayed in the
IC SCPI Monitoring window.
Communication has not
been established from
this window.
Establish connection by
clicking the Connect to IC
button.
Communication has not
been established on the
GPIB or Ethernet port.
Ensure that all cables are
properly connected.
Ensure that the Remote
option is selected and
that all parameters are
properly set.
This message is displayed on
Windows desktop: RAID
Volume Degraded.
Malfunctioning hard
drive.
Contact EXFO.
Problem
Probable Cause
Recommended Action