beautypg.com

Product-specific commands-description, Product-specific commands—description, Calibration[1..n]:zero – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 96

background image

SCPI Command Reference

90

IQS-3150

Product-Specific Commands—Description

Product-Specific Commands—Description

:CALibration[1..n]:ZERO

Description

Returns the mechanism of the instrument to its

home position.

This command is an event and has no associated

*RST condition or query form.

Syntax

:CALibration[1..n]:ZERO

Parameter(s)

None

Example(s)

STAT:QUES:BIT9:COND? Returns 1 if the

instrument's mechanism should be returned to

its home position.

STAT? Must return READY

CAL:ZERO

STAT:OPER:BIT9:COND? Keep resending the

query as long as the operation is not complete

(returned value is not 0).

Notes

This command will take at least 15 seconds to

complete.

See Also

:SENSe[1..n]:CORRection:COLLect:ZERO

:STATus:OPERation:BIT:CONDition?

:STATus:QUEStionable:BIT:CONDition?