SCPI Command Reference

98

IQS-3150

Product-Specific Commands—Description

:INPut[1..n]:OFFSet

Description

This command sets an offset value for the

attenuation. This offset value will be added to the

absolute attenuation. The same offset value will

be used for all wavelengths. This value is used

only when the ATTenuation control mode is

active. The offset is only taken into account when

the INPut[1..n]:RATTenuation command is used.

At *RST, the current value is set to 0 dB.

Syntax

:INPut[1..n]:OFFSetDB]

>|MAXimum|MINimum|DEFault

Parameter(s)

Offset:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is DB. The special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

highest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter corresponds to a valid

offset value. You can use the INPut[1..n]:OFFSet?

MAX and INPut[1..n]:OFFSet? MIN queries to

determine a valid range for the attenuation

offset.