Input[1..n]:reference – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual
Page 113

SCPI Command Reference
Variable Attenuator
107
Product-Specific Commands—Description
:INPut[1..n]:REFerence
Description
This command sets, for the current wavelength,
a reference value for the attenuation. When the
instrument is used in Reference mode, the
attenuation configured with the
INPut[1..n]:RATTenuation command is relative to
this reference value. This command will have no
effect when the instrument is used in Absolute or
X + B mode. This value it used only when the
ATTenuation control mode is active.
At *RST, the reference value that will be set
depends on the instrument you have.
Syntax
:INPut[1..n]:REFerence [ Parameter(s) Reference: defined as a followed by an optional DATA> element. The allowed PROGRAM DATA> element is DB. The MAXimum and DEFault are accepted on input. MINimum allows to set the instrument to the lowest supported value. MAXimum allows to set the instrument to the highest supported value. DEFault allows the instrument to select a value for the The reference value.
The program data syntax for