SCPI Command Reference

116

IQS-3150

Product-Specific Commands—Description

:OUTPut[1..n]:ALC[:STATe]?

Description

This query indicates if the power tracking that

controls the output power level has been

activated or not. The power tracking is done via

the ALC (Automatic Leveling Control) loop.

At *RST, this value is set to off.

Syntax

:OUTPut[1..n]:ALC[:STATe]?

Parameter(s)

None

Response Syntax

Response(s)

State:
The response data syntax for is defined

as a

element.

The response corresponds to the state

of the ALC (Automatic Leveling Control) loop.

0: No power tracking will be performed.

1: Power tracking is activated.

See Also

:OUTPut[1..n]:ALC[:STATe]
:CONTrol[1..n]:MODE

:OUTPut[1..n]:DTOlerance

:OUTPut[1..n]:POWer

:OUTPut[1..n]:RPOWer