Introducing the IQS-3150 Variable Attenuator

Variable Attenuator

7

Main Software Features

Output power

Multiple Module Control

The multimodule application is used to control and monitor several
IQS-3150 variable attenuators at the same time.

Display Mode

Description

Absolute

(default mode)

The displayed output power takes into account both the absolute

output power value and the offset value.

Displayed power = absolute power + offset value

Reference

The displayed output power value is relative to a defined reference

value (see Selecting a Display Mode on page 30).

Displayed power = absolute power – reference value + offset value

X+B

The displayed output power is the sum of two elements:

³

X = output power of the unit.

³

B = correction factor. You must define the correction factor for the

test wavelength before enabling the X+B mode (refer to Setting
the B Value
on page 23).

Displayed power = absolute power + correction factor

a

+ offset value

a.

The correction factor value is specific to the wavelength.