:OUTPut[1..n]:ALC[:STATe]

Description

This command activates or deactivates the

power tracking that controls the output power

level. The power tracking is done via the ALC

(Automatic Leveling Control) loop. The state of

the ALC loop (on or off) is used only when the

POWer control mode is active.

At *RST, this value is set to off.

Syntax

:OUTPut[1..n]:ALC[:STATe]

Parameter(s)

State:
The program data syntax for is defined

as a element. The

special forms ON and OFF are accepted

on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.

The parameter corresponds to the new

state of the ALC (Automatic Leveling Control)

loop.

OFF: No power tracking will be performed.

ON: Power tracking is activated.

See Also

:OUTPut[1..n]:ALC[:STATe]?

:CONTrol[1..n]:MODE

:OUTPut[1..n]:DTOlerance

:OUTPut[1..n]:POWer

:OUTPut[1..n]:RPOWer