beautypg.com

Control modes (self-adjusting modules) – EXFO IQS-3150 Variable Attenuator for IQS-500/600 User Manual

Page 11

background image

Introducing the IQS-3150 Variable Attenuator

Variable Attenuator

5

Main Software Features

Control Modes (Self-Adjusting Modules)

With the self-adjusting modules, you can select between two control
modes: Attenuation and Output Power.

Control Mode

Description

Attenuation

The instrument applies the attenuation values you have selected.
The value appearing at the top of the data display (dB) represents

your attenuation setting whether it was entered in a test sequence or

on the Instrument function tab.
The value appearing underneath the attenuation value is the output

power.

Output Power

The instrument produces the output power level you have selected.
The value appearing at the top of the data display (dBm) represents

your power level setting whether it was entered in a test sequence or

on the Instrument function tab.
The value appearing underneath the attenuation value is the output

power.